Advanced Foot Energizer General Specifications
| Remote Control Unit | |
| Type of wireless technology | Infrared ray (Wavelength: 940nm) |
| Operating frequencies | 455KHz |
| Operating voltage | 3V |
| Maximum output powers | 0.1W |
| Operating distances and ranges | 15-20m |
| Intended use environment | Home |
| Functions | Turn on/off the device; select mode and the intensity of output. |
| Basic Unit Characteristics | |
| Power Source | Adapter Input: 100-240Vac, 50-60Hz, 0.1A
Output: 5Vdc, 1A Unit Input: 5Vdc, 1A |
| Method of Line Current Isolation | Type BF Applied Part |
| Patient leakage current | Comply with IEC 60601-1 and IEC 60601-2-10 |
| Normal Condition | AC: 54.50μA DC: 0.5μA |
| Single Fault Condition | AC:120.00μA, DC: 0.6μA |
| Average DC current through electrodes when device is on but no pulses are being applied | <0.01μA |
| Number of channels | 2 |
| Number of modes | 25 |
| Output Intensity Level | 99 steps |
| Output Intensity | 0V~112V |
| Indicator Display | Indicate the following information: Power on/off, intensity level, selected mode, time remaining of an application mode. |
| Time Range (minutes) | 25 |
| Main Unit Weight | 2kg |
| Weight of Accessories | Adapter: 275g, Electrode :40g, Remote Control: 20g |
| Main Unit Dimension | 428mm x 428.8mm x 185mm |
| Electrode Pads Size | Foot: Conductive Silicone Rubber* (PP conductive plastic): 254mm × 98mm × 13mm, 99.27 cm2
Small: 88mm × 58mm × 3.2mm, 50.04 cm2 |
| Electrode Wires Length | 1785±2mm |
| AC Adapter Wire Length | 1445±2mm |
| Housing Materials of Main Unit | ABS plastic |
| Basic Unit Characteristics Continued | ||
| Accessories Materials | Conductive Foot Pad Material: PP (PP conductive plastic)
Red Transparent Plastic: ABS Plastic Small Electrode Pads: white silica gel, black conductive silicone, transparent conductive adhesive silicone, transparent PET silicone AC Adapter Housing: ABS plastic Remote Controller: ABS plastic Sole Massage Roller: ABS plastic |
|
| Output Specification | ||
| Waveform and Shape | Pulsed, symmetric biphasic, rectangular | |
| Maximum Output Voltage (+/- 10%) | 44V @ 500Ω | 80V @ 2KΩ | 112V @ 10KΩ | |
| Maximum Output Current (+/- 10%) | 88mA @ 500Ω | 40mA @ 2KΩ | 11.2mA @ 10KΩ | |
| Net Charge (per pulse) | 0μC @ 500Ω | |
| Maximum Phase Charge | 12.78μC @ 500Ω | |
| Maximum Average Current | 0.968mA @ 500Ω | |
| Maximum Current Density | 0.235mA/cm² @ 500Ω | |
| Maximum Power Density | 1.38mW/cm² @ 500Ω | |
| Pulse Duration | 120μs | |
| Pulse Cycle | 12.94ms | |
| Frequency | 77.3Hz | |
| Contraction and Relaxation time | Adjustable, due to different modes.
Contraction time: 200ms ~ 56s Relaxation time: 0~0.5s |
|
| Treatment Time | 25 min (default) | |
| ON Time | 0.6 s | |
| OFF Time | 0.6 s | |
| Additional Features | |
| Environment for Operation | Temperature: 5 ~ 45° C
Humidity: 20 ~ 65% RH |
| Environment for Storage | Temperature: 0 ~ 45° C,
Humidity: 10 ~ 90% RH Electrode Pad: 10~20° C |
*Technically, silicone is considered part of the rubber family. But, if you define plastics widely, silicone is something of a hybrid between a synthetic rubber and a synthetic plastic polymer.